Skip to content

Peptas Instruments / Preparative HPLC system

Peptas Instruments HPP-1004 Preparative HPLC with UV detection, 600 mL/min class

Preparative HPLC system

Made to order. Quote on request.

Request a quotation Explore specifications
Generic technical illustration of a modular preparative chromatography system.
Illustrative render — shown for illustration only.

Key features

Technical detail

Specifications

Values are shown with their stated basis and source.

HPP-1004 technical specifications
ParameterSpecificationBasis / conditionsSource
Flow-rate range0.01–599.99 mL/minNominal; Selected column of the configuration table. Configuration-specificSource 1 · Configuration table / infusion unit / flow rate / selected column
Flow output error±1 %Nominal; The source cites a national test standard without stating test conditions. Configuration-specificSource 1 · Configuration table / infusion unit / flow output error / selected column
Flow stability0.5 %Maximum; The source cites a national test standard without stating test conditions; not relabelled as RSD. Configuration-specificSource 1 · Configuration table / infusion unit / flow stability / selected column
System pressure limit15 MPaMaximum; System limit; no pressure-versus-flow envelope is stated. Configuration-specificSource 1 · Configuration table / infusion unit / system pressure / selected column
Pressure pulsation0.5 MPaMaximum; Test conditions are not stated by the source. Configuration-specificSource 1 · Configuration table / infusion unit / pressure pulsation / selected column
Pump-head material316L as standard; PEEK, PCTFE or titanium alloy optionalNominal; Pump-head material only; not a statement about every wetted component. Configuration-specificSource 1 · Configuration table / note below the table / pump head
DetectorUV detector as standard; single- or dual-wavelength mode selectableNominal; Standard detector type. Configuration-specificSource 1 · Configuration table / detection unit / wavelength range and note below the table
UV wavelength range190–400 nmNominal; Selected UV range; the extended-range variant is not included. Configuration-specificSource 1 · Configuration table / detection unit / wavelength range
LampDeuterium; tungsten lamp optionalNominal; Configuration-specificSource 1 · Configuration table / detection unit / light source
Wavelength accuracy±1 nmNominal; Test conditions are not stated by the source. Configuration-specificSource 1 · Configuration table / detection unit / wavelength accuracy
Wavelength precision±0.1 nmNominal; Test conditions are not stated by the source. Configuration-specificSource 1 · Configuration table / detection unit / wavelength precision
Detector absorbance range0–3 AUNominal; Stated detector range. Configuration-specificSource 1 · Configuration table / detection unit / absorbance range
Column inner diameter100 mmNominal; Separation unit of this configuration, as stated by the source. Configuration-specificSource 1 · Configuration table / separation unit / inner diameter / selected column
Recommended column typeDynamic axial compression columnsNominal; Recommendation as stated by the source. Configuration-specificSource 1 · Configuration table / separation unit / recommended column type / selected column
Sample introductionSample loading by pumpNominal; Configuration-specificSource 1 · Configuration table / sample unit / selected column
Fraction collection (optional unit)Collection by retention time, by response value or manually, among other modesNominal; Optional fraction collection unit; not standard equipment. Configuration-specificSource 1 · Configuration table / optional fraction collection unit
Specification sources
  1. Specification source 1Document SHA-256: d57a5e86590e90853176ed288d57c329f66647fb1e2292542e2493e715b285b3

Source documents are supplied with the quotation on request.